Philips provides update on Philips Respironics’ PE-PUR sound
abatement foam test and research program
June 28, 2022
Testing in accordance with ISO 18562 and ISO 10993 standards
conducted by five certified, independent testing laboratories in
the US and Europe
Amsterdam, the Netherlands – On June 14, 2021,
Royal Philips’ (NYSE: PHG; AEX: PHIA) subsidiary Philips
Respironics, initiated a voluntary recall notification/field safety
notice* to address potential health risks related to the
polyester-based polyurethane (PE-PUR) sound abatement foam in
specific CPAP, BiPAP and mechanical ventilator devices.
At the time the recall notification/field safety notice* was
issued, Philips Respironics relied on an initial, limited data set
and toxicological risk assessment, and assumed a worst-case
scenario for the possible health risks out of an abundance of
caution. Since then, together with five certified, independent
testing laboratories in the US and Europe, as well as other
qualified third-party experts, Philips Respironics has been
conducting a comprehensive test and research program on the PE-PUR
foam to better assess and scope the potential patient health risks
related to possible emission of particulates from degraded foam and
volatile organic compounds (VOCs).
This update is intended to provide healthcare providers,
patients, and other stakeholders with updated information on the
testing results to date. Philips will continue to provide regular
updates as new test results and assessments become available, as
not all tests have been completed to date.
The overall guidance for healthcare providers and patients in
the recall notification/field safety notice* remains unchanged
at this time.
Testing methodsTesting results and conclusions
to date are grouped by device air path design and configuration,
i.e., based on how the air flows through the device. Of the five
device categories, the first-generation DreamStation devices
represent 68% of the registered affected devices globally. Within
each device category, testing and analyses are performed on new
devices with pristine foam, devices with lab-aged foam and used
devices:
- Visual assessment of the foam in returned/used devices to
assess the prevalence of visible foam degradation.
- VOC testing to identify and quantify organic compounds that may
be inhaled during device use.
- Particulate Matter (PM) testing to determine concentrations of
respirable particulates, i.e., particulates up to 10
micrometers in diameter, as it relates to inhalation risks and
established health thresholds.
- Additional physical, chemical and biological testing of the
PE-PUR foam related to patient risks if patients were in contact
with foam material.
The complete update on the PE-PUR testing results and
conclusions available to date can be found here, and the main
findings have been presented below. Healthcare providers, patients,
and other stakeholders should use the complete update for any
informed decision making, and not the overview in this press
release.
“I deeply regret the concern experienced by patients who rely on
the affected Respironics sleep and respiratory devices for their
health and quality of life and want to emphasize our commitment to
providing them with a resolution as fast as possible,” said Frans
van Houten, CEO of Royal Philips. “More than 1,000 of our
colleagues are working extremely hard to achieve this. While
certain prolonged tests across the affected product categories are
still to be completed, the results to date for the first-generation
DreamStation devices, which represent the majority of the
registered affected devices, show a very low prevalence of visible
foam degradation. In addition, the new and used first-generation
DreamStation devices passed volatile organic compound and
respirable particulate emission testing. This is very encouraging.
Results to date also indicate that ozone cleaning significantly
exacerbates foam degradation.”
First-generation DreamStation devices (68% of
registered affected devices globally)Visual
assessment: In order to determine the prevalence of foam
degradation, which may result in foam particulate emission, a
visual assessment of the foam was performed on a sample of 60,847
returned/used first-generation DreamStation devices from the US and
Canada. The visual inspection was conducted according to a specific
protocol as part of the repair process. The sample included 36,341
devices for which the users reported no use of ozone cleaning,
11,309 devices for which the users reported use of ozone cleaning,
and 13,197 devices for which it was reported unknown by the user
whether ozone cleaning was used.
- Prevalence of visible foam degradation (US and
CA): 164 of 36,341 (0.5%) devices with self-reported no
ozone use showed significant visible foam degradation.
- Impact of repeated ozone cleaning: Devices
with self-reported ozone use were 14x more likely to have
significant visible foam degradation than those with self-reported
no ozone use: 777 of 11,309 devices (7%) showed significant visible
foam degradation.
- Assessment of a sample of devices linked to reported
alleged foam degradation complaints: 422 devices of the
inspected 60,847 returned/used devices are linked to a reported
foam degradation complaint. However, only 18 out of these 422
devices (4%) actually showed visible foam degradation.
In those devices where visible foam degradation was significant,
i.e., there was reduction in foam volume, it was observed that
there was accumulation of degraded foam within the airpath inside
the device. The foam becomes hygroscopic (i.e., absorbs moisture)
and sticky with degradation. It also loses significant volume and
increases density as the structure changes from a foam to a viscous
liquid material. As such, even when foam particulates are formed by
degradation, they are likely to accumulate within the device and
may not be directly emitted by the device.
A visual assessment of the foam was also performed on a sample
of 1,360 returned/used first-generation DreamStation devices from
various countries in Europe and on a sample of 931 returned/used
devices from Japan.
- Prevalence of visible foam degradation (EU and
JPN): None of the assessed devices from Europe or Japan
showed significant visible degradation.
Volatile Organic Compounds (VOCs): VOC testing according to ISO
18562-3 was performed on new, lab-aged and used devices to (1)
quantify VOC emissions from devices, and (2) assess the
toxicological risk associated with exposure to the quantified
concentrations of those VOCs. It is important to note that these
tested new and lab aged first-generation DreamStation devices were
not exposed to ozone cleaning, in accordance with the instructions
for use.
- As previously provided in an update on December 23, 2021,
VOC emissions are below established limits based
on ISO 18562-3 testing and evaluation of new, lab-aged, and used
devices. Exposure to the level of VOCs identified to date for the
first-generation DreamStation devices is not anticipated to
result in long-term health consequences for patients.
Particulate Matter (PM): PM testing according to ISO 18562-2 was
performed on the devices to (1) quantify the particulate matter
emitted from devices, and (2) assess whether the concentration
detected is less than thresholds provided in the
standard.
- New devices and used devices (including used devices with
visible foam degradation) were tested and were all found to
be compliant with ISO 18562-2 allowable limits for
PM emissions.
- Tested PM emissions of used devices with degradation were not
statistically different than PM emissions without degradation,
suggesting that degradation did not contribute to
appreciable elevated levels of respirable particles in the devices
tested.
- The used devices that were tested for PM emissions, were also
evaluated for cleanliness based on a visual inspection of the
exterior of the device, i.e., the presence of environmental
materials on the external surface of the device, such as the inlet
filter location. For these devices, average particulate
matter counts in devices classified as ‘dirty’ were significantly
greater than those classified as ‘clean’.
Biocompatibility testing of (degraded) PE-PUR foam: Additional
testing is still being performed in accordance with ISO 10993 to
facilitate a toxicological risk assessment of (degraded) foam
particulates, which is relevant if they would potentially contact
the patient. This testing includes chemical characterization (i.e.,
what chemicals may potentially extract or leach from the foam and
have direct contact with body tissues and/or fluids), in vitro
assessment (i.e., tests performed in a test tube, dish, etc.
outside the body), and in vivo assessment (i.e., preclinical
testing) of new, lab aged and/or used PE-PUR foam.
To support the assessment of potential genotoxicity,
cytotoxicity, and irritation risks of lab-aged foam, chemical
characterization of lab-aged foam, as well as experiments to assess
the amount of foam that can potentially contact the patient are
being conducted, as lab-aged foam did not pass Ames (genotoxicity),
cytotoxicity and skin irritation bioassay testing. Further,
complementing the lab-aged foam assessment, a chemical
characterization of returned/used degraded foam is also being
conducted to better elucidate risks under field conditions. Per ISO
10993, the bioassay results cannot stand alone and, therefore, a
positive Ames, cytotoxicity, or skin irritation result triggers a
required follow-up evaluation including identification of potential
confounding factors, and a weight of evidence assessment to
determine a confirmed conclusion on potential risks for patients
under expected usage of the device.
Other devices under the recall notification/field safety
notice Other devices that are being tested include
DreamStation Go (1% of the registered devices) and SystemOne (26%
of the registered devices). These devices each have a different air
path design/configuration compared to the first-generation
DreamStation devices but contain the same PE-PUR foam.
New DreamStation Go and SystemOne devices passed VOC and
PM testing based on standards available prior to ISO 18562
i.e., Indoor Air Quality Evaluation (as previously disclosed in the
April 25, 2022, update). Further ISO 18562 VOC and PM testing is
ongoing. The results of ISO 10993 biocompatibility testing of
degraded PE-PUR foam for the first-generation DreamStation devices,
which is ongoing, will also apply to DreamStation Go and SystemOne
devices.
Further devices include Trilogy 100/200 (3% of the registered
devices) and OmniLab/A-Series BiPAP (2% of the registered devices).
New Trilogy 100/200 passed VOC and PM testing.
New OmniLab devices passed VOC and PM testing based on
standards available prior to ISO 18562 i.e., Indoor Air
Quality Evaluation (as previously disclosed in the April 25, 2022,
update). New and used OmniLab devices passed VOC testing based on
ISO 18562. Further testing is ongoing.
Summary of ongoing testsThe first-generation
DreamStation, DreamStation Go and SystemOne CPAP/BiPAP devices
represent 95% of the registered affected devices globally. Philips
Respironics expects to complete the remaining VOC and PM testing
for these devices, as well as the degraded foam toxicological risk
assessments (in accordance with ISO 10993) in the coming months.
Philips Respironics will also continue with the tests to assess the
impact of repeated ozone cleaning on foam degradation in these
CPAP/BiPAP devices, as well as the remaining VOC and PM testing and
the degraded foam toxicological risk assessments for the Trilogy
100/200 and OmniLab ventilator devices.Silicone foam
testing per FDA’s November 2021 requestIn November 2021,
the FDA requested that Philips retain an independent laboratory to
perform additional testing to determine what, if any, potential
safety risks may be posed to patients by silicone-based foam.
Philips Respironics engaged independent testing laboratories to
perform additional VOC testing. Based on the draft reports, Philips
Respironics has not identified any safety issues. The
assessment is being completed, and the final reports are subject to
FDA review, which are expected in the coming months.
Guidance for healthcare providers and
patientsAs indicated, this update is intended to provide
healthcare providers, patients, and other stakeholders with updated
information on the testing results to date. The overall
guidance for healthcare providers and patients in the
recall notification/field safety
notice* remains unchanged at this
time. Philips Respironics remains fully committed to
addressing all devices affected by the recall notification/field
safety notice* and continues to work with the relevant competent
authorities to further optimize the remediation plan. To date,
approximately 2.7 million replacement devices and repair kits have
been produced.
Additional informationFurther information,
including the complete update and FAQs, as well as video messages
from Philips CEO Frans van Houten, Chief Business Leader Connected
Care Roy Jakobs and Technical Project Manager for the test and
research program Jan Bennik, can be found here and here.
* Voluntary recall notification in the US/field
safety notice for the rest of the world.
For media questions, please
contact:
Steve KlinkPhilips Global Press OfficeTel.: +31 6
10888824E-mail: steve.klink@philips.com
Ben ZwirsPhilips Global Press OfficeTel.: +31 6 15213446E-mail:
ben.zwirs@philips.com
About Royal PhilipsRoyal Philips (NYSE: PHG,
AEX: PHIA) is a leading health technology company focused on
improving people's health and well-being, and enabling better
outcomes across the health continuum – from healthy living and
prevention, to diagnosis, treatment and home care. Philips
leverages advanced technology and deep clinical and consumer
insights to deliver integrated solutions. Headquartered in the
Netherlands, the company is a leader in diagnostic imaging,
image-guided therapy, patient monitoring and health informatics, as
well as in consumer health and home care. Philips generated 2021
sales of EUR 17.2 billion and employs approximately 78,000
employees with sales and services in more than 100 countries. News
about Philips can be found at www.philips.com/newscenter.
Forward-looking statements This statement
contains certain forward-looking statements with respect to the
financial condition, results of operations and business of Philips
and certain of the plans and objectives of Philips with respect to
these items. Examples of forward-looking statements include
statements made about the strategy, estimates of sales growth,
future EBITA, future developments in Philips’ organic business and
the completion of acquisitions and divestments. By their nature,
these statements involve risk and uncertainty because they relate
to future events and circumstances and there are many factors that
could cause actual results and developments to differ materially
from those expressed or implied by these statements.
Koninklijke Philips NV (EU:PHIA)
Historical Stock Chart
From Feb 2023 to Mar 2023
Koninklijke Philips NV (EU:PHIA)
Historical Stock Chart
From Mar 2022 to Mar 2023